Test C
hSelect up to five large, slower-drying stems from a section
of the windrow.
hPlace them one at a time across two adjacent points on the
# 831 short pin prod.
hThe average of these stem readings should be about two to
five points higher than the actual moisture content.
Notes
hRepeat these steps in different parts of the field and pay
special attention to the areas where the hay is heaviest.
hThe amount of variation found among windrow readings as
well as the average stem moisture should be taken into
consideration before the decision is made to start baling.
FACTORS AFFECTING YOUR
READINGS
Because of the many variables that affect the electrical meter
readings, the indicated moisture content should not be used
as an absolute quantitative measurement. Meter readings are
very useful guidelines for the safe storability of hay.
Meter readings become more significant when they are
considered in the light of the density of the bales, anticipated
handling and storage, and prevailing climate conditions.
Range of Moisture Content
The F-2000 is designed to test moisture in hay over a range
of 6%-40%. Readings over 30% should be used only as a
qualitative indication of high moisture content. Delmhorst
moisture meters use the relationship existing between electrical
conductivity and moisture content in hay. As moisture content
increases, so does the conductivity.
Tests on hay at high moisture content, over 25%, are less
accurate. This is mostly due to the variability in moisture
distribution. The reduced level of accuracy in the high range
does not significantly affect the usefulness of the meter, as a
few high readings indicate that some action be taken to dry
the hay to avoid spoilage or even self-combustion.
While it is important to note the average of several readings,
it is even more important to note the high readings and the